Since 2000 · 3,000+ customers

Yokogawa · EJA110E differential pressure transmitter

Yokogawa EJA110E spares and repair in India

The Yokogawa EJA110E is a standard traditional-mount differential pressure transmitter from the EJA-E series, used for flow, level and pressure measurement. Star Automations sources EJA110E amplifier boards, capsule assemblies, indicators and covers, and repairs amplifier boards at board level where circuits and parts allow. Send the full model code from the nameplate and we confirm availability or repairability.

Star Automations is an independent multi-brand repair and supply company and is not an OEM service centre.

At a glance

Where the EJA110E parts sit, and how an enquiry works

Diagram of a differential pressure transmitter on an orifice-plate flow line, with its five main parts numbered Flow 12345 Differential pressure transmitter on an orifice-plate flow line
Typical arrangement; your installation may differ.
  1. Integral indicator
  2. Amplifier (CPU) assembly
  3. Terminal block and cover
  4. Capsule (sensor) assembly
  5. Flange gasket and bolts
  1. STEP 1
    Send the model codeA nameplate or label photo on WhatsApp is enough.
  2. STEP 2
    We checkAvailability and repairability, before you ship anything.
  3. STEP 3
    Supply, repair or alternativeThe route that is practical for your part.
  4. STEP 4
    Test and returnEach repaired unit is bench-tested first.
01The transmitter

What the EJA110E is, in plain terms

Per Yokogawa’s published data, the EJA110E measures differential pressure as its main variable and can also report static pressure. It is used for liquid, gas and steam flow (with an orifice plate or similar primary element), liquid level, density and pressure.

ItemPublished data
SeriesEJA-E, DPharp sensor technology, traditional-mount body
Reference accuracy±0.055% of span for differential pressure (0.04% option)
Rangeability100:1
Response time90 ms (differential pressure)
Capsule codesF, M, H and V, each covering a different span range
Safety ratingYokogawa lists SIL 2 for a single transmitter and SIL 3 for a dual-transmitter configuration
Communication4-20 mA output with BRAIN, HART, FOUNDATION Fieldbus or PROFIBUS PA, depending on the output code in the model string

Figures are Yokogawa’s published data for a standard unit and vary with the exact model code. Check your own nameplate and the Yokogawa general specification for your suffix codes.

02Parts

EJA110E parts people ask for

This list follows typical failure patterns. Your fault may differ, so send the symptom with the model code.

PartWhy it is usually neededWhat we do
Amplifier (CPU) assemblyOutput stuck or erratic, no HART or BRAIN communication, blank displaySupply, or board-level repair where circuits and parts allow
Capsule (sensor) assemblyReading drift or no response after overpressure, corrosion or process-side damageSupply of a matching capsule, or advice to replace the transmitter
Integral indicatorCracked or blank displaySupply
Cover, O-ring and terminal blockMoisture ingress, damaged cover threads, corroded terminalsSupply
Process flange gasket and boltsLeakage at the cover flange after serviceSupply on request
03Quote details

Read the full model code before you order

Two EJA110E units can differ in capsule range, wetted-part material, output signal and approvals. A capsule or amplifier that fits one unit may not suit another. The nameplate carries what we need.

  • Full MODEL line, for example EJA110E followed by the suffix and option codes, including anything after the slash
  • Capsule code (F, M, H or V) and the calibrated range
  • Output signal and communication type
  • STYLE code and serial number
  • A photo of the nameplate and the symptom in one line
04Before you replace

Check the loop first, then the transmitter

Many “transmitter faults” are in the impulse lines or the wiring. These checks take a few minutes with your plant’s own procedure and save a wrong purchase.

SymptomCheck firstLikely part if the transmitter is at fault
Reading stuck or noisyBlocked or leaking impulse line, manifold valve position, air or condensate in the lineCapsule or amplifier
No HART or BRAIN communicationLoop resistance, supply voltage at the terminals, wiring polarityAmplifier assembly
Blank display, loop current presentIndicator connector and cover contactIntegral indicator
Zero shift after overpressure or cleaningRe-trim zero with the transmitter isolated; compare with a referenceCapsule assembly
Water in the terminal compartmentCover seal, conduit entry and cable glandTerminal block, O-ring, possibly the amplifier

Follow your site permit and isolation procedure and Yokogawa’s instruction manual when working on a live loop.

05Options

Supply, repair or replace

Supply

New surplus, refurbished-and-tested or used parts, as the part allows. The condition is stated in the quote.

Repair

Amplifier boards are repaired at board level where circuits and parts allow. We confirm repairability from your photo before you ship, and test the unit before return.

Replace

If the capsule is damaged or the model is discontinued, we tell you plainly whether a matching part or a replacement transmitter is the practical route. See repair or replace advice.

07Enquiry

Send the model code

Name, contact and the nameplate photo are enough to start.

We use these details to reply to your enquiry. See our privacy policy.

08FAQ

Common questions

What is the Yokogawa EJA110E?

It is a standard traditional-mount differential pressure transmitter in Yokogawa’s EJA-E series. Plants use it for flow measurement with an orifice plate, for liquid level and for pressure. It reports differential pressure and can also report static pressure.

Do you supply EJA110E spare parts in India?

Yes. We source EJA110E amplifier assemblies, capsule assemblies, indicators, covers and terminal parts as new surplus, refurbished-and-tested or used stock. The condition is stated in the quote. Send the full model code and we confirm availability.

Can you repair an EJA110E amplifier board?

We repair amplifier boards at board level where circuits and parts allow, and confirm repairability from the nameplate and a photo before you ship. A damaged capsule usually means a replacement capsule or transmitter rather than a repair, and we tell you which is practical.

What should I send for a quote?

The full model line with suffix and option codes, the capsule code and range, the style and serial number, a photo of the nameplate, the symptom, the quantity and the delivery location. WhatsApp +91 97866 22233 works well.

Is an EJX110A the same as an EJA110E?

No. The EJX110A is a related model from the earlier EJX series. Parts are not always interchangeable between the two, so confirm with the full model code before ordering.

How do I choose between repairing and replacing the transmitter?

Repair usually makes sense when the amplifier or indicator has failed and the capsule is healthy. Replacement is usually the better route when the capsule is damaged, the unit is badly corroded or the model is discontinued. Send the fault and nameplate and we give an honest recommendation.

Are you a Yokogawa authorised dealer or service centre?

No. Star Automations is an independent multi-brand repair and supply company and is not an OEM service centre. For a new transmitter under OEM warranty, contact Yokogawa or its channel partners.

Is a repaired transmitter suitable for a safety-instrumented loop?

Treat a repaired or non-OEM transmitter as unsuitable for a certified SIL function unless your functional-safety assessment accepts it. The safety rating applies to the Yokogawa-certified configuration. For safety loops, ask your safety engineer before reusing any repaired unit.